Point-of-need detection of microcystin-LR using a smartphone-controlled electrochemical analyzer
Tian Guan,Wenzheng Huang,Ningxia Xu,Zhenzhen Xu,Lelun Jiang,Mengting Li,Xiaoqun Wei,Yingju Liu,Xing Shen,Xiangmei Li,Changqing Yi,Hongtao Lei
DOI: https://doi.org/10.1016/j.snb.2019.05.028
2019-09-01
Abstract:<p>The development of rapid and <em>in situ</em> detection method for microcystin-LR (MC-LR) with portable, reliable, and easy to use devices are highly demanded, but it is still deficient. In this study, a portable and wireless point-of-need detection system for microcystin-LR (MC-LR) quantitation was developed and validated. The whole system consisted of three main parts: screen printed carbon electrodes (SPCEs), smartphone-controlled electrochemical analyzer (SCEA), and Android's smartphone. The SCEA was a homemade ARM STM32 microcontroller-based electrochemical analyzer, which have similar performance compared with a commercially available electrochemical workstation. A smartphone installed with a user-friendly application was employed to control the analyzer, receive and analyze data, and display detection results in real-time. This system can accurately quantitate MC-LR in the range of 0.001–100 µg/L with a detection limit of 0.00011 µg/L, and the results were consistent with LC–MS/MS. In addition, a preliminary MC-LR contamination map was generated using this smartphone-based platform by measuring water samples at 6 reservoirs located in Pearl River Delta (Guangdong Province, China). The cost-effective, reliable and easy-to-use smartphone-based detection system reported here can facilitate active toxicant screening to guarantee the safety of the drinking water, particularly in resource-limited area.</p>