Liver-Buda-Sage: Simultaneous Whole Liver T2 and T*2 Mapping in One Breath-Hold.

Zijing Zhang,Huihui Ye,Maosen Li,Huafeng Liu,Berkin Bilgic
DOI: https://doi.org/10.1109/isbi52829.2022.9761635
2022-01-01
Abstract:We propose liver-BUDA-SAGE (liver Blip Up-Down Acquisition for Spin And Gradient Echo imaging), an efficient echo-planar imaging (EPI) sequence to rapidly obtain high-resolution T 2 and $T_2^{\ast}$ quantitative mapping of the human liver at 3.0 Tesla (T) with high-coverage and high geometric fidelity. This multi-shot navigator-free acquisition includes multiple $T_{{2^ - }}^{\ast},T_{{2^ - }}^\prime$ and T 2 -weighted contrasts with alternating phase-encoding polarities across the interleaved shots. A field map estimated from interim reconstructions was incorporated into the joint multi-shot EPI reconstruction with a structured low-rank constraint to eliminate geometric distortion. An adapted quantitative multi-parametric mapping method was established using a combined least-square solution and Bloch dictionary matching to obtain maps based on the re-constructed echoes. In-vivo results demonstrate the ability of liver-BUDA-SAGE to provide high-coverage, distortion-free, multi-contrast images and quantitative ${T_2}/T_2^{\ast}$ maps at 1.5×1.5 mm 2 in-plane resolution in ~20s for 16 slices. Estimated quantitative maps showed good agreement in phantom compared to conventional mapping methods with slight biases of 2.6 ms and -0.2 ms for T 2 and $T_2^{\ast}$. Measured in-vivo liver values (mean±SD) for T 2 and $T_2^{\ast}$ were 21.26±2.37 ms and 30.59±3.70 ms which are consistent with those reported in the literature.