A Flexible Terahertz Metamaterial Biosensor for Cancer Cell Growth and Migration Detection

Weihao Fang,Xiaoqing Lv,Zhengtai Ma,Jian Liu,Weihua Pei,Zhaoxin Geng
DOI: https://doi.org/10.3390/mi13040631
IF: 3.4
2022-01-01
Micromachines
Abstract:Metamaterial biosensors have been extensively used to identify cell types and detect concentrations of tumor biomarkers. However, the methods for in situ and non-destruction measurement of cell migration, which plays a key role in tumor progression and metastasis, are highly desirable. Therefore, a flexible terahertz metamaterial biosensor based on parylene C substrate was proposed for label-free and non-destructive detection of breast cancer cell growth and migration. The maximum resonance peak frequency shift achieved 183.2 GHz when breast cancer cell MDA-MB-231 was cultured onto the surface of the metamaterial biosensor for 72 h. A designed polydimethylsiloxane (PDMS) barrier sheet was applied to detect the cell growth rate which was quantified as 14.9 mu m/h. The experimental peak shift expressed a linear relationship with the covered area and a quadratic relationship with the distance, which was consistent with simulation results. Additionally, the cell migration indicated that the transform growth factor-beta (TGF-beta) promoted the cancer cell migration. The terahertz metamaterial biosensor shows great potential for the investigation of cell biology in the future.
What problem does this paper attempt to address?