Noninvasive Blood Pressure Measurement Based on Photoplethysmography

Yudong Wu,Shuncong Zhong
DOI: https://doi.org/10.1007/978-981-10-4109-9_13
2017-01-01
Abstract:Stationary wavelet transform (SWT) and the Artificial Neural Networks (ANN) Algorithm were proposed for continuous measurement of blood pressure. A total number of 26,900 photoplethysmography (PPG) signals from Mimic Database were analyzed using SWT decomposition. The fifth layer high frequency coefficient was employed to reconstruct the signals. Ten characteristic parameters obtained from the reconstructed signals, as ANN input vector, and the PPG corresponding to the systolic blood pressure (SBP) and diastolic blood pressure (DBP), as the ANN vector output, were used to train the ANN model. The averaging error between the systolic pressure estimated by the proposed method and actual pressure value is 3.98 mmHg, and the standard deviation is 3.32 mmHg. The averaging error between the diastolic pressure estimated by the proposed method and actual pressure value is 3.81 mmHg, and the standard deviation is 3.63 mmHg. These results satisfy the American National Standards of the Association for the Advancement of Medical Instrumentation.