Rapid Identification of Bacterial Mixtures in Urine Using MALDI-TOF MS-based Algorithm Profiling Coupled with Magnetic Enrichment

Wenmin Cheng,Haimei Shi,Mengjing Teng,Menghuan Yu,Bin Feng,Chuanfan Ding,Shaoning Yu,Fan Yang
DOI: https://doi.org/10.1039/d1an02098f
2021-01-01
The Analyst
Abstract:Urinary tract infections (UTIs) are a severe public health problem caused by mono- or poly-bacteria. Culture-based methods are routinely used for the diagnosis of UTIs in clinical practice, but those are time consuming. Rapid and unambiguous identification of each pathogen in UTIs can have a significant impact on timely diagnoses and precise treatment. Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) is an alternative method for the identification of pathogens in clinical laboratories. However, a certain number of pure bacteria are required for MALDI-TOF MS analysis. Here, we explored a strategy combining magnetic enrichment and MALDI-TOF MS for the rapid identification of pathogenic bacterial mixtures in urine. Fragment crystallizable mannose-binding lectin-modified Fe3O4 (Fc-MBL@Fe3O4) was used for rapid enrichment and the individual-peak-based similarity model as the analytical tool. Within 30 min, a mixture of the four most prevalent UTI-causing bacteria, Escherichia coli, Klebsiella pneumoniae, Staphylococcus aureus, and Pseudomonas aeruginosa, was successfully identified using this method. This rapid MALDI-TOF MS-based strategy has potential applications in the clinical identification of UTI pathogens.
What problem does this paper attempt to address?