The Intracavitary ECG Method for Positioning the Tip of Central Venous access Devices in Pediatric Patients: Results of an Italian Multicenter Study
Francesca Rossetti,Mauro Pittiruti,Massimo Lamperti,Ugo Graziano,Davide Celentano,Giuseppe Capozzoli
DOI: https://doi.org/10.5301/jva.5000281
2014-02-09
The Journal of Vascular Access
Abstract:Purpose The Italian Group for Venous Access Devices (GAVeCeLT) has carried out a multicenter study investigating the safety and accuracy of intracavitary electrocardiography (IC-ECG) in pediatric patients. Methods We enrolled 309 patients (age 1 month-18 years) candidate to different central venous access devices ( VAD) - 56 peripherally inserted central catheters (PICC), 178 short term centrally inserted central catheters (CICC), 65 long term VADs, 10 VADs for dialysis - in five Italian Hospitals. Three age groups were considered: A (<4 years, n = 157), B (4-11 years, n = 119), and C (12-18 years, n = 31). IC-ECG was applicable in 307 cases. The increase of the P wave on IC-ECG was detected in all cases but two. The tip of the catheter was positioned at the cavo-atrial junction (CAJ) (i.e., at the maximal height of the P wave on IC-ECG) and the position was checked during the procedure by fluoroscopy or chest x-ray, considering the CAJ at 1-2 cm (group A), 1.5-3 cm (group B), or 2-4 cm (group C) below the carina. Results There were no complications related to IC-ECG. The overall match between IC-ECG and x-ray was 95.8% (96.2% in group A, 95% in group B, and 96.8% in group C). In 95 cases, the IC-ECG was performed with a dedicated ECG monitor, specifically designed for IC-ECG (Nautilus, Romedex): in this group, the match between IC-ECG and x-ray was 98.8%. Conclusions We conclude that the IC-ECG method is safe and accurate in the pediatric patients. The applicability of the method is 99.4% and its feasibility is 99.4%. The accuracy is 95.8% and even higher (98.8%) when using a dedicated ECG monitor.