Evaluation of the performance of systems for whole blood C-reactive protein detection: a multi-center study

李怀远,陈红兵,杨治理,张泓,潘秋辉,奎莉越,莫丽亚,柯江维,徐锦,罗红权,向赟,唐振华,渠巍,程娟,田礼军,吕欣,吴亦栋,李贵霞,王雨新,郑磊,李向阳,张文利,叶辉铭,刘海鹏,马丽娟,佘尚扬,杨俊梅,贾卉
DOI: https://doi.org/10.3760/cma.j.cn114452-20210219-00112
2021-01-01
Abstract:Objective:To explore the performance of the commonly used whole blood C-reactive protein (CRP) detection systems and give related recommendation on the performance requirements of detection systems.Methods:A total of 7 540 venous blood samples from 26 maternal, child and children′s hospitals were collected to conduct this multi-center study on the analytical performance of 5 commonly used whole blood CRP detection systems from March to April in 2019. The blank check, carryover, repeatability, intermediate precision, linearity, sample stability, influence of hematocrit/triglyceride/bilirubin, comparison with SIEMENS specific protein analyzer and trueness were evaluated. The 5 systems included BC-5390CRP autohematology analyzer, AstepPLUS specific protein analyzer, Ottoman-1000 Automated Specific Protein POCT Workstation, i-CHROMA Immunofluorometer equipment Reader and Orion QuikRead go detecting instrument. The 5 systems were labeled as a, b, c, d and e randomly.Results:Within the 5 systems, all values of blank check were less than 1.00 mg/L, the carryovers were lower than 1.00%. The repeatability of different ranges of CRP concentrations including 3.00-10.00, 10.00-30.00 and>30.00 mg/L were less than 10.00%, 6.00% and 5.00%, respectively, and the intermediate precision was less than 10.00%. The linearity correlation coefficients of the 5 systems were all above 0.975, while the slope was within 0.950-1.050. Whole blood samples were stable within 72 hours both at room temperature (18-25 ℃) and refrigerated temperature (2-8 ℃). The CRP results were rarely influenced by high triglyceride or bilirubin, except for the immmunoturbidimetric test based on microparticles coated with anti-human CRP F(ab) 2 fragments. When triglyceride was less than 15.46 mmol/L, the deviation of CRP was less than 10.00%. When bilirubin was less than 345.47 μmol/L, the deviation of CRP was less than 10.00%. CRP was more susceptible to Hct on the systems without Hct correction. The deviation of CRP between different Hct dilution concentration and 40% dilution concentration can reach as high as 67.48%. The correlation coefficients ( r) of 5 systems were all more than 0.975 in the range of 0-300.00 mg/L compared with Siemens specific protein analyzer. All systems passed the trueness verification using the samples with specified values of 12.89 and 30.60 mg/L. Conclusion:The performance of 5 systems can basically meet the clinical needs, but it is suggested that the whole blood CRP detection system without automatic Hct correction should be modified manually.
What problem does this paper attempt to address?