Left ventricular hemodynamics with veno‐arterial extracorporeal membrane oxygenation
Rajat Kalra,Tamas Alexy,Jason A. Bartos,Anthony R. Prisco,Marinos Kosmopoulos,Valmiki R. Maharaj,Alejandra Gutierrez Bernal,Andrea M. Elliott,Santiago Garcia,Ganesh Raveendran,Ranjit John,Daniel Burkhoff,Demetris Yannopoulos
DOI: https://doi.org/10.1002/ccd.30951
IF: 2.3
2024-01-12
Catheterization and Cardiovascular Interventions
Abstract:Background There is considerable debate about the hemodynamic effects of veno‐arterial extracorporeal membrane oxygenation (VA‐ECMO). Aims To evaluate the changes in left ventricular (LV) function, volumes, and work in patients treated with VA‐ECMO using invasive LV catheterization and three‐dimensional echocardiographic volumes. Methods Patients on VA‐ECMO underwent invasive hemodynamic evaluation due to concerns regarding candidacy for decannulation. Hemodynamic parameters were reported as means±standard deviations or medians (interquartile ranges) after evaluating for normality. Paired comparisons were done to evaluate hemodynamics at the baseline (highest) and lowest tolerated levels of VA‐ECMO support. Results Twenty patients aged 52.3 ± 15.8 years were included. All patients received VA‐ECMO for refractory cardiogenic shock (5/20 SCAI stage D, 15/20 SCAI stage E). At 3.0 (2.0, 4.0) days after VA‐ECMO cannulation, the baseline LV ejection fraction was 20% (15%, 27%). The baseline and lowest VA‐ECMO flows were 4.0 ± 0.6 and 1.5 ± 0.6 L/min, respectively. Compared to the lowest flow, full VA‐ECMO support reduced LV end‐diastolic volume [109 ± 81 versus 134 ± 93 mL, p = 0.001], LV end‐diastolic pressure (14 ± 9 vs. 19 ± 9 mmHg, p
cardiac & cardiovascular systems