16-Channel Readout Circuit for Nucleic Acid Detection Based on Graphene Electrolyte-gated Field-effect Transistors (Egfets)

Wei Zhang,Yunlin Liu,Zhibo Chen,Yating Zou,Yizhou Jiang,Ye Lu,Yajie Qin
DOI: https://doi.org/10.1109/asicon52560.2021.9620258
2021-01-01
Abstract:This paper proposed a readout circuit for nucleic acid detection based on the graphene electrolyte gated field effect transistors (EGFETs). The nucleic acid sequence can be inferred by detecting the channel current of the EGFET when it works. The readout circuit allows a large detection range which covers 20pA - 10μA. In order to detect weak signal, baseline elimination part is set in the circuit. The range of baseline elimination is 2μA - 200μA. The readout circuit is designed with 0.18μm CMOS process. The power consumption is 44.53μW when the sensitivity is set to be 20pA.