Heteroatom-free conjugated tetraphenylethylene polymers for selectively fluorescent detection of tetracycline

Qiu Zhuang,Cheng Zhang,Haiyan Zhuang,Hanyu Deng,Xiangpeng Lin,Yan Li,Hao Chen,Aming Xie,Wei Dong
DOI: https://doi.org/10.1016/j.aca.2021.339236
IF: 6.911
2022-01-01
Analytica Chimica Acta
Abstract:The antibiotic tetracycline (Tc) is a major contaminant in food and water, with adverse effects on both ecosystems and human health. The development of novel sensors for tetracycline detection is of great importance. In this work, we develop a novel heteroatom-free conjugated tetraphenylethylene polymer (TPE-CMP) fluorescence sensor for the detection of tetracycline. In the presence of Tc, the emission fluorescence of TPE-CMP was quenched by the photoinduced electron transfer mechanism to achieve high sensitivity. The polymers can detect tetracycline at a concentration of 0-100 mu g/mL with a good linear correlation (0.99), and the limit of detection (LOD) is 1.23 mu g/mL. Furthermore, TPE-CMP has excellent selectivity in detecting Tc in the presence of various anti-interference analytes, including ions and antibiotics. In addition, the practical feasibilities of TPE-CMP for Tc sensing were further investigated in milk, urine and wastewater samples with satisfactory recoveries (from 94.96% to 112.53% for milk, from 96.41% to 99.31% for urine and from 98.54% to 100.52% for wastewater). We have designed and synthesized TPE-CMP based on heteroatom-free for the specific fluorescence detection of tetracycline, expanding the range of fluorescence detection sensors and offering great promise for practical applications. (C) 2021 Published by Elsevier B.V.
What problem does this paper attempt to address?