Rapid Determination Of Ultra-Trace Plutonium In Large Volume Of Urine Samples By Tandem Quadrupole Inductively Coupled Plasma-Mass Spectrometry

Yang Wu,Yihong Xu,Shan Xing,Xiongxin Dai,Ni Yuan,Maoyi Luo
DOI: https://doi.org/10.1016/j.sab.2021.106259
2021-01-01
Abstract:Rapid bioassay method for plutonium in urine samples are employed to assess the possible internal dose resulting from exposure to plutonium in the event of a radiological and nuclear emergency. In this work, a rapid method for the determination of plutonium isotopes (Pu-239 and Pu-240) in large volume of urine samples was developed using tandem quadrupole ICP-MS combined with a high-efficiency sample introduction system (APEX-Omega). The mixture gases of NH3 and He was employed to effectively eliminate the tailing of U-238 and the uranium hydride ions ((UH+)-U-238-H-1 and (UH2+)-U-238-H-1) and significantly improve the measurement sensitivity of plutonium. The method was validated by analyzing urine samples spiked with known amounts of( )(239)Pu and Pu-240 ranging from 1 to 60 fg. The detection limits for the method were estimated to be 0.83 fg for Pu-239, 0.14 fg for Pu-240 in 1 L of urine.
What problem does this paper attempt to address?