The Sensitivity And Specificity Of Ultrasonic Localization Of Endotracheal Intubation

Yongtao Yang,Yun Yue,Dongya Zhang
2017-01-01
Abstract:Objective: This study aimed to evaluate the application value of portable ultrasound on clinical judgment of endotracheal tube placement. Methods: This was a blinded prospective study. A total of 94 ASAI similar to III patients in Beijing Chaoyan hospital from September, 2013 to February 2014, were randomly selected. All patients were intended to perform routine orotracheal intubation under general anesthesia. The tracheal intubations with laryngoscopy were performed by residents, 10 mL of saline were injected up to the endotracheal tube cuff (COVIDIENTM. MallinckrodtTM. Taper Guard Evac Oral Tracheal Tube). The endotracheal tube placement (in the trachea or in the esophagus) was confirmed by ultrasound scanning of the neck. End tidal CO2 (ETCO2) wave combined with chest ausculation were used as the gold standard of successful tracheal intubation. Sensitivity, specificity, positive predictive value and negative predictive value of endotracheal tube placement were calculated. Software SPSS 17.0 was used for statistical analysis. Results: Ninety-three patients were enrolled (ASAI similar to III) in this study with 42 males and 51 females. The mean age of the 93 patients was 53.5 +/- 11.8 years old; BMI was 24.91 +/- 3.38. Esophageal intubation was detected in 9 patients. One case of esophageal intubation was misdiagnosed by sonography. The sensitivity and specificity of ultrasonic localization of tracheal intubations was 100% (95% confidence interval, 94.6%-100%) and 88.9% (95% confidence interval, 84%-100%) with the positive and negative predictive value of 98.8% (95% CI 92.7-99.9) and 100.0% (95% CI 59.8-100) respectively. Conclusion: Sonography is portable, non-invasive and repeatable method; it could be used to confirm intracheal intubation with a relatively high sensitivity and specificity.
What problem does this paper attempt to address?