A Novel Amperometric Nitric Oxide Sensor Based on Polythionine / Nation Modified Glassy Carbon Electrode

XX Chen,F Wang,YX Xu,SS Hu
2006-01-01
Abstract:A novel amperometric sensor for the determination of nitric oxide was developed by coating polythionine/nafion on a glassy carbon electrode. This sensor exhibited a great enhancement to the oxidation of nitric oxide. The oxidation peak currents were linear to the concentration of nitric oxide over the wide range from 3.6×10^(-7) to 6.8×10^(-5)mol.L^(-1), and the detection limit was 7.2×l0^(-8)mol.L^(-1). Experimental results showed that this nitric oxide sensor possessed excellent selectivity and longer stability. NO releasing from rat kidney was monitored by this sensor.
What problem does this paper attempt to address?