Clinical Application Of Roentgen Stereophotogrammetric Analysis Technique To Detect Hip Aseptic Loosening
Jin-Wu Wang,Yuan-Jing Xu,Han Yang,Ke-Rong Dai
DOI: https://doi.org/10.1007/978-981-13-6924-7_9
2019-01-01
Abstract:Artificial joint replacement is a common treatment for joints affected by trauma, osteoarthritis, or rheumatoid arthritis. Prosthetic aseptic loosening is a major clinical problem and one of the most important reasons for revisions after prosthetic replacement. Routine clinical testing to detect aseptic loosening involves physical and imaging examinations; however, they are not sufficiently accurate. The gold standard of orthodontic kinematics measurement is the Roentgen stereophotogrammetric analysis (RSA) technique, which can predict aseptic loosening of implants by measuring micromovements during the first 2 years after joint replacement. The accuracy of RSA technique is in the range 0.05-0.5 mm in translation and 0.15-1.15 degrees in rotation. The RSA technique plays an important role in evaluating implant fixations with high precision and three-dimensional (3D) migration measurements. This chapter explains the basic principles and development of the RSA technique as well as its clinical application to detect hip artificial joint aseptic loosening.