Direct and Simultaneous Determination of Methionine Sulfoxide and Pyroglutamic Acid Impurities in Compound Amino Acid Injection‐18 AA by Ion‐pair Reversed‐phase HPLC

Li,Yaru Qin,Bingyuan Ma,Hongyan Cui,Ping Qiao,He Jiang,Ming Zhang
DOI: https://doi.org/10.1002/sscp.201800059
2018-01-01
Separation Science Plus
Abstract:A direct and simultaneous determination of methionine sulfoxide and pyroglutamic acid impurities in Compound Amino Acid Injection-18 AA by ion-pair reversed-phase high-performance liquid chromatography with ultraviolet detection is described. The separation of both impurities in Compound Amino Acid Injection-18 AA was accomplished with gradient elution consisting of acetonitrile and 10 mM diammonium hydrogen phosphate buffer (containing 15 mM sodium 1-heptanesulfonate and obtaining pH 2.3 by addition of phosphoric acid). The limits of detection (mu g/mL, S/N = 3:1) and quantification (mu g/mL, S/N = 10:1) of methionine sulfoxide and pyroglutamic acid were 0.02 and 0.08, and 0.06 and 0.19, respectively. The linearity was in the range of 0.08-200.20 mu g/mL for methionine sulfoxide and 0.19-201.20 mu g/ml- for pyroglutamic acid. The relatively simple method proved accurate (recovery 99.68-102.98% for methionine sulfoxide, 98.10-102.00% for pyroglutamic acid, n = 9) and precise (repeatability RSD = 1.06% for methionine sulfoxide, RSD = 0.60% for pyroglutamic acid, n = 6) in its application to commercial Compound Amino Acid Injection-18 AA.
What problem does this paper attempt to address?