Potentiometric Detection of Trace-Level Chlorpyrifos in Seawater Using a Polymeric Membrane Electrode Coupled with On-Line Molecularly Imprinted Solid-Phase Extraction

Sixia Li,Rongning Liang,Wei Qin,Ruiqing Yao
DOI: https://doi.org/10.1016/s1452-3981(23)05080-0
IF: 1.541
2015-01-01
International Journal of Electrochemical Science
Abstract:Potentiometric sensors have evolved to be a promising tool for environmental trace analysis and potentiometric biosensing. However, the applications of potentiometric sensors are restricted to measurements of samples without complex matrixes. A new potentiometric detection system for detection of trace-level chlorpyrifos (CPF) in real samples is described in this paper. By using on-line molecularly imprinted solid-phase extraction (MISPE), interferences from sample matrixes can be effectively eliminated. Numerous key variables including affecting the extraction recovery of MISPE have been optimized. The proposed detection system offers a low detection limit of 0.027 nmol L-1 and exhibits excellent selectivity over other organicphosphate pesticides such as parathion-methyl, phoxim and dipterex. The practical application of the proposed system has been carried out for detection of CPF at trace levels in real seawater samples.
What problem does this paper attempt to address?