Numerical Analysis Of Middle Ear Pressure Effects On Performance Of Implantable Hearing Devices

Jiabin Tian,Jing Zhang,Zhushi Rao,Na Ta,Lifu Xu
2015-01-01
Abstract:Middle ear pressure has a significant influence on the dynamic behavior of human ear. To analyze the effect of this physiological condition on the implant performance of implantable hearing devices, a finite element model of human ear consisting of the external ear canal, middle ear and passive cochlea was first developed through Micro-CT imaging and reverse engineering technology. Validation of this model was accomplished by comparing the dynamic responses with experimental data in the literature. Then the middle ear stiffening due to static pressure varying from -2.0 to +2.0 kPa was considered by altering the elastic modulus of components in middle ear. Based on this model, a piezoelectric actuator was further designed and attached to the incus long process. By means of acoustic-structure-fluid and electro-mechanical coupled analysis, equivalent sound pressures (ESP) of the actuator under different static pressure levels were finally calculated. The results show that ESPs of the actuator were decreased with the load of middle ear pressure in the frequency range of 250 to 2500 Hz, and increased at frequencies of 2500-8000 Hz. Meanwhile, negative pressure had a greater influence on the actuator performance than positive pressure, especially at frequency below 2500 Hz. This work will benefit the design and evaluation of implantable middle-ear hearing devices for special physiological conditions.
What problem does this paper attempt to address?