A Simple and Fast LC-MS/MS Method for the Simultaneous Determination of Tenofovir Alafenamide and Tenofovir in Human Plasma

Xiaoping Qian,Qi Chen,Yunhui Chen,Shiliang Ji,Ying Wang,Ye Sun,Huixin Qi,Kan Zhong,Jinfang Jiang,Xin Chen,Lifeng Huang,Yong Yang
DOI: https://doi.org/10.1002/bmc.5273
2021-01-01
Biomedical Chromatography
Abstract:A simple and fast liquid chromatography-tandem mass spectrometry method was established and validated for the simultaneous determination of tenofovir alafenamide (TAF) and tenofovir (TNF) in human plasma. A simple protein precipitation procedure was employed to extract analytes from plasma. Chromatographic separation was performed on an Eclipse Plus C-18 column utilizing a fast gradient elution starting with 2% of 2 mM ammonium acetate-formic acid (100/0.1, v/v) followed by increasing the percentage of acetonitrile. Detection was performed on a tandem mass spectrometer equipped with an electrospray ionization source operated in the positive ionization mode, using the transitions m/z 477.2 -> m/z 346.1 for TAF and m/z 288.1 -> m/z 176.1 for TNF. TAF-d(5) and TNF-d(7) were used as the internal standard of TAF and TNF, respectively. The method was validated in the concentration ranges 1.25-500 ng/mlfor TAF and 0.300-15.0 ng/ml for TNF with acceptable accuracy and precision.
What problem does this paper attempt to address?