A Needle Tip CCEA Microfluidic Device Based on Enhanced Dean Flow for Cell Washing

Xin Shi,Wei Tan,Yuwen Lu,Wenfeng Cao,Guorui Zhu
DOI: https://doi.org/10.1038/s41378-021-00311-9
2021-01-01
Abstract:Particle/cell washing is an essential technique in biological and clinical manipulations. Herein, we propose a novel circular contraction-expansion array (CCEA) microdevice. It can be directly connected to a needle tip without connection tubes. Its small size and centrosymmetric structure are beneficial to low sample consumption, high connection stability, and a wide application range. Computational fluid dynamics (CFD) simulation results show that the CCEA structure can produce a stronger Dean flow and lead to faster particle/cell focusing than the circle structure and CEA structure with the same length. Experimentally, an optimal flow rate ratio of 1:3 and an optimal total flow rate of 120 mu L/min were found to ensure a stable fluid distribution. Under these conditions, rapid focusing of 10-20 mu m particles with high efficiencies was achieved. Compared with a normal CEA device using tubes, the particle loss rate could be reduced from 64 to 7% when washing 500 mu L of a rare sample. Cell suspensions with concentrations from 3 x 10(5)/mL to 1 x 10(3)/mL were tested. The high cell collection efficiency (>85% for three cell lines) and stable waste removal efficiency (>80%) reflected the universality of the CCEA microfluidic device. After the washing, the cell activities of H1299 cells and MCF-7 cells were calculated to be 93.8 and 97.5%, respectively. This needle-tip CCEA microfluidic device showed potential in basic medical research and clinical diagnosis.