Clinical applicability of optical surface imaging system in thorax tumor radiation therapy
Cheng-wei YE,Lu SU,Qiang SHANG,Bin DU,Yan-long WU,Sen BAI,Ren-ming ZHONG
DOI: https://doi.org/10.13339/j.cnki.sglc.20180517.009
2018-01-01
Abstract:Objective To evaluate the clinical applicability of the optical surface imaging (OSI) system in thorax tumor radiation therapy. Methods From February to October in 2015, 11 patients (113 radiation therapy fractions) with thorax tumor radiation therapy were enrolled, which included 8 males (89 radiation therapy fractions) and 3 females (24 radiation therapy fractions), aged 46-77 years old with mean age of 59 years old. Among 11 patients, 6 were primary lung cancer, 3 were metastatic lung cancer and 2 were esophageal cancer. All of them were fixed with heat plastic in supine position, and hands crossed with elbows on forehead. Before the first treatment, the cone beam computed tomography (CBCT) scan was given and corrected the displacement error, then OSI system was used to acquire body surface image as the reference image. From the second time, corrected the set-up error with the reference image of OSI system, then CBCT scan was given, and the set-up error data of each scan in 2 systems were recorded and analyzed. Results The results of translational error in CBCT registration x (left and right), y (head and foot), z (front and back), and rotation directions rx (PIT), ry (ROL), and rz (ROT) was (-0.83 ± 3.15) mm, (-1.40 ± 3.89) mm, (0.71 ± 2.08) mm and (0.29 ± 0.76) °, (0.36 ± 0.81) °, (0.10 ± 1.00) °, respectively; OSI was (0.01 ± 1.34) mm, (-0.12 ± 1.26) mm, (-0.26 ± 1.19) mm and (-0.14 ± 0.57) °, (0.10 ± 0.60) °, (-0.25 ± 0.69) °, respectively. Subtraction two system, the result was (-0.85 ± 3.42) mm, (-1.29 ± 3.80) mm, (0.98 ± 2.38) mm, (0.42 ± 0.98) °, (0.26 ± 1.16) ° and (0.36 ± 1.24) °, respectively (P <0.05). The standard deviations after subtraction translation errors in x, y and z was 3.42 mm, 3.80 mm and 2.38 mm, respectively. The rotation difference between 2 systems was small, and the standard deviation about 1°. At x, y and z of translational process in radiotherapy, the intra-fractional error detected by c-motion of CatalystTM system was (-0.17 ± 1.25) mm, (0.47 ±1.42) mm and (-0.17 ± 1.96) mm, respectively, and (-0.04 ± 0.51) °, (0.13 ± 0.42) °, (-0.19 ± 0.76) ° for rx, ry and rz rotations, and maximum deviation value was (2.54 ± 1.10) mm in each direction. Conclusion It is demonstrated that OSI system and CBCT system showed certain differences in detection and correction of chest tumor radiation therapy set-up error, and OSI system could monitor the intra-fractional error in real-time. As the set-up error of chest radiotherapy is affected by many factors, the comprehensive utilization strategy of OSI system and CBCT system needs to be further studied.