Accelerated K‐space Shift Calibration for Free‐breathing Stack‐of‐radial MRI Quantification of Liver Fat And

Xiaodong Zhong,Tess Armstrong,Chang Gao,Marcel D. Nickel,Fei Han,Brian M. Dale,Xinzhou Li,Sevgi G. Kafali,Peng Hu,Holden H. Wu,Vibhas Deshpande
DOI: https://doi.org/10.1002/mrm.28981
2021-01-01
Magnetic Resonance in Medicine
Abstract:Purpose To develop an accelerated k-space shift calibration method for free-breathing 3D stack-of-radial MRI quantification of liver proton-density fat fraction (PDFF) and R2*. Methods Accelerated k-space shift calibration was developed to partially skip acquisition of k-space shift data in the through-plane direction then interpolate in processing, as well as to reduce the in-plane averages. A multi-echo stack-of-radial sequence with the baseline calibration was evaluated on a phantom versus vendor-provided reference-standard PDFF and R2* values at 1.5T, and in 13 healthy subjects and 5 clinical subjects at 3T with respect to reference-standard breath-hold Cartesian acquisitions. PDFF and R2* maps were calculated with different calibration acceleration factors offline and compared to reference-standard values using Bland-Altman analysis. Bias and uncertainty were evaluated using normal distribution and Bayesian probability of difference (P < .05 considered significant). Results Bland-Altman plots of phantom and in vivo data showed that substantial acceleration was highly feasible in both through-plane and in-plane directions. Compared to the baseline calibration without acceleration, Bayesian analysis revealed no significant differences on biases and uncertainties of PDFF and R2* measurements with all acceleration methods in this study, except the method with through-plane acceleration equaling slices and averages equaling 20 for PDFF and R2* (both P < .001) for the phantom. A six-fold reduction in equivalent calibration acquisition time (time saving >= 25 s and >= 80.7%) was achieved using recommended acceleration factors for the in vivo protocols in this study. Conclusion This proposed method may allow accelerated calibration for free-breathing stack-of-radial MRI PDFF and R2* mapping.
What problem does this paper attempt to address?