A macro–micro FE and ANN framework to assess site-specific bone ingrowth around the porous beaded-coated implant: an example with BOX® tibial implant for total ankle replacement

Minku,Rajesh Ghosh
DOI: https://doi.org/10.1007/s11517-024-03034-x
2024-02-09
Medical & Biological Engineering & Computing
Abstract:The use of mechanoregulatory schemes based on finite element (FE) analysis for the evaluation of bone ingrowth around porous surfaces is a viable approach but requires significant computational time and effort. The aim of this study is to develop a combined macro–micro FE and artificial neural network (ANN) framework for rapid and accurate prediction of the site-specific bone ingrowth around the porous beaded-coated tibial implant for total ankle replacement (TAR). A macroscale FE model of the implanted tibia was developed based on CT data. Subsequently, a microscale FE model of the implant-bone interface was created for performing bone ingrowth simulations using mechanoregulatory algorithms. An ANN was trained for rapid and accurate prediction of bone ingrowth. The results predicted by ANN are well comparable to FE-predicted results. Predicted site-specific bone ingrowth using ANN around the implant ranges from 43.04 to 98.24%, with a mean bone ingrowth of around 74.24%. Results suggested that the central region exhibited the highest bone ingrowth, which is also well corroborated with the recent explanted study on BOX®. The proposed methodology has the potential to simulate bone ingrowth rapidly and effectively at any given site over any implant surface.
engineering, biomedical,computer science, interdisciplinary applications,mathematical & computational biology,medical informatics
What problem does this paper attempt to address?