Preparation of pencil-lead bismuth-film electrode and its determination of Pb2+ in drinking water

Miaomiao Gu,Fuyang Wang,Aidong Zhou
DOI: https://doi.org/10.16791/j.cnki.sjg.2017.11.020
2017-01-01
Abstract:A pencil-lead bismuth-film electrode is constructed by introducing a pencil-lead electrode and combing the bismuth with high resolution and good sensitivity .The Pb2+ is detected by using a pencil core bismuth electrode based on the anodic stripping voltammetry . The experimental conditions are optimized , and the stripping voltammogram with good peak shape stripping signal is obtained . The pencil-lead bismuth-film electrode shows a good linear relationship with the Pb2+ response ,the linear range is between 4—10 μg/L and 15—85 μg/L ,and the detection limit of Pb2+ reaches 0 .46 μg/L .The detection results meet the drinking water requirements of the European Union ,United States Environmental Protection Agency and World Health Organization .
What problem does this paper attempt to address?