Abstract:Deteriorated carbon fiber microelectrodes was regenerated by graphene oxide using electrodeposition method, and the electrochemical behaviors of dopamine(DA)solution on the regenerated electrode was investigated by cyclic voltammetry(CV)and differential pulse voltammetry(DPV). The electrodepositing time and voltage of graphene oxide were also optimized. The results showed that in pH 7.2, 20 mmol/L Tris-HCl buffer solution, the regenerated electrode showed excellent electrochemical performance for DA and NE. Under the optimized condition, the concentration of DA showed good linear relationships with the oxidation peak current in the range of 0.1~ 100 μmol/L, with the limits of detection of 0.1 μmol/L.