Determination the effective silicon in soil by inductively coupled plasma emission spectra (ICP-OES)

Daying ZHOU,Xiaojian GONG,Hongbo ZHONG
DOI: https://doi.org/10.16614/j.gznuj.zrb.2018.03.009
2018-01-01
Abstract:The method of determining effective silicon content in soil was established by ICP-OES.The effects of ICP-OES working conditions and spectrometric lines on the effective silicon detection results were studied.The results showed that the detected limit of this method was low (0.03μg/mL),with the higher precision of this method (0.25% ~ 0.92%),and the recovery rate of this method was good (95.9% ~ 104.5%).The results of the test and the national soil effective state level were consistent with the standard material certificate.This determination method was high sensitivity,good repeatability,high accuracy and feasible operation,which was suitable for the determination of effective silicon in soil.
What problem does this paper attempt to address?