Temperature Rise of Electromagnet of Large-Gap Magnetic Drive System

Yan Xu,Jian-ping Tan,Jun-feng Wang,Yun-long Liu,Zhong-yan Zhu
DOI: https://doi.org/10.3969/j.issn.1000-565X.2013.11.018
2013-01-01
Abstract:In order to guarantee stable and reliable operation of micro axial-flow blood pumps,the temperature rise of electromagnet in the large-gap magnetic drive system should be investigated and accurately controlled.In this paper,the temperature distribution,heat loss,heat flux and thermal gradient of various components of the system were analyzed,and a simulation model of the thermal field of the electromagnet was established.Then,the distri-bution and variation law of the system temperature field were obtained via finite element simulation and experiment investigation.Finally,to meet the requirements of human body adaptability and long-term running of the electro-magnet,the electromagnetic and environmental parameters of the electromagnetic drive system were simulated and optimized by taking a suitable temperature rise of the electromagnet as the target.The results show that the tempera-ture rise of the electromagnet can be effectively reduced by decreasing the loss and improving the heat emission con-dition of the electromagnet;and that the highest temperature of the electromagnet does not exceed 50℃ when the current of electromagnet coils is lower than 0.56 A and when the air velocity is higher than 3.5 m/s.
What problem does this paper attempt to address?