Research on identification of bacteria in bloodstream infections based on FT-NIR characteristic spectrum selection

Jianming WANG,Wu WANG,Xianghui LI,Yurong LI
DOI: https://doi.org/10.7631/issn.1000-2243.2017.05.0713
2017-01-01
Abstract:Fourier transform near-infrared spectroscopy (FT-NIR) was used to collect the range from 1 000 to 1 852 nm near-infrared spectra of the three common pathogen bacteria which included Escherichia coli,Pseudomonas aeruginosa,and Staphylococcus aureus.Competitive adaptive reweighted sampling (CARS) algorithm to be used to select the characteristic wavelength variables which were used to build the PLS-DA model and LS-SVM model.Moreover,performance of PLS-DA and LS-SVM model built by the characteristic wavelength variables were also compared with PLS-DA and LS-SVM model built by full spectra.Studies showed that full spectra model had poor prediction performance due to some wavelength variables contains irrelevant information.The prediction performance of PLS-DA model and LS-SVM had improved significantly by using 24 characteristic wavelengths selected by CARS.Meanwhile,LS-SVM model achieved the optimal performance which the correct rate of three pathogen bacteria was 85.0%,100%,100% respectively.The results showed that CARS can remove useless information,reducing model complexity,enhanced model prediction performance and combined with LS-SVM can accurately identify clinical pathogens.