Linear Measurement and Correction for the 106 Dynamic Range of the FY-3 Solar Backscattered UV Ozone Spectrometer

YANG Xiaohu,WANG Shurong,HUANG Yu
DOI: https://doi.org/10.3788/m0002120123801.0019
2012-01-01
Optical Technique
Abstract:Based on the requirement of linear measurement to the FY-3 solar backscattered UV ozone spectrometer(SBUS) which has a dynamic range of 106,a way with neutral density filter method and double optical stop method combined together is proposed.This new idea is in a position to the linear measurement in the whole dynamic range to the FY-3 solar backscattered UV ozone spectrometer.The nonlinearity is corrected subsequently.Uncertainty analysis shows combined standard uncertainty of 0.44% to this new method and 0.62% to the traditional law of distance square method for a range of 103-104,which indicates effectively the validation and superiority of this new idea for the large dynamic range of the photodetectors.