High Performance Liquid Chromatographic Determination of Thyroxine in Iodinated Casein with Barium Hydroxide Method

王宗义,常碧影,张丽英,杨文军,董涛
DOI: https://doi.org/10.3969/j.issn.1004-4957.2005.01.023
2005-01-01
Abstract:A high performance liquid chromatographic method was developed for the determination of thyroxˉine(T 4 )in iodinated casein(IC).The IC was hydrolyzed with bariumhydroxde solution.It was found that the relative highest stable yield of T 4 from IC was obtained under the following conditions:barium octahydrate amount added into per milliliter of IC solution ranged from0.5to1.0g;the hydrolyzing time at110℃ranged from8to12hours;and the pH range was controlled within12to13after precipitating the bariumions with sulphuric acid.The separation was performed on a Waters Symmetry shield TM RP18column(5μm,3.9mm×150mm).T 4 could be separated completely using phase A(water,0.1%formic acid)and phase B(acetonitrile,0.1%formic acid)as mobile phase at a flowrate of0.8mL/min,and10%to50%of phase B as eluant for40min.The T 4 was then quantitatively determined by external standard method at240nm.A good linear calibration curve was obtained for0.5to25mg·L -1 T 4 ,with a detection limit of0.2mg·g -1 .
What problem does this paper attempt to address?