Quantification Of Methimazole In Plasma And Tablet Samples By Surface Enhanced Raman Spectroscopy In Combination With Multiplicative Effects Model

Hu Min,Chen Zeng-Ping,Chen Yao,Shi Cai-Xia,Yu Ru-Qin
DOI: https://doi.org/10.11895/j.issn.0253-3820.141086
IF: 1.193
2015-01-01
Chinese Journal of Analytical Chemistry
Abstract:The surface-enhanced Raman spectroscopy (SERS) technique in combination with multiplicative effects model (MEMSERS) was applied to the quantification of methimazole in plasma and tablet samples. Experimental results showed that by using g-mercapto-isonicotinic acid as internal standard and nanosilver particles as enhanced substrate, the MEMSERS model can be applied to the accurate quantitative determination of methimazole in plasma and tablet samples-The average relative predictive error for the determination of methimazole in plasma samples was 5.1% . The recovery rates of MEMSERS for methimazole in tablet samples were in the range of 93.3% - 110.9%, comparable to the corresponding values of LC-MS/MS experiments-The limit of detection was estimated to be 32 nmol/L.
What problem does this paper attempt to address?