Abstract:Ebola glycoprotein was captured by immunomagnetic microspheres,followed by formation of an immune sandwich complex with biotinylated antibody.Then,3,3',5,5'-tetramethylbenzidine (TMB) was developed by streptavidin-horseradish peroxidase (SA-HRP) to form a colored product,which had a strong absorption at 370 nm.Finally,Ebola glycoprotein was quantified with the absorbance at 370 nm.Immunomagnetic microspheres was characterized by immune fluorescence.Reliability of detection principle was proved by control experiments.At the optimized conditions,the absorbance was linear with the concentration of Ebola glycoprotein in the range of 1.0-25.0 ng/mL,with a detection limit of 0.18 ng/mL.This method showed good reproducibility,high specificity and nice capacity of resisting disturbance.