The Determination of Total Kjeldahl Nitrogen in Plants by AA3 Flow Analyzer
Li-ping ZHANG,Jiu-rong WANG,Hong-zhao YUAN,Mei-mei GENG,Chun-yong LI,Wen CHEN,Can PENG,Li-wei XU
DOI: https://doi.org/10.16498/j.cnki.hnnykx.2016.010.024
2016-01-01
Abstract:Total Kjeldahl nitrogen in plants was determined by the AA3 flow analyzer, the NaOH weight of the buffer in different acidity samples, and linear range, repeatability, precision, recovery rate and the detection limit were investigated. Compared with the analysis results measured by AA3 and FIAstar 5000, the results were consistent. Using 0.10 and 0.23 mL/min sample tube, the detection range was 0-70 and 0-50 mg/L respectively. The correlation coefifcient of the standard curve was more than 0.999, the average recovery rate was 97.10%-105.30%, and the detection limit was 0.07-0.08 mg/L. The results showed that the method was accuracy and could be widely used.