Performance Parameter Test of IRFPA Used in Astronomy Imaging Detection

冯志伟,杨潇,白先勇,段帷,宋谦,吴宏,张志勇
DOI: https://doi.org/10.7517/j.issn.1674-0475.2017.02.140
2017-01-01
Abstract:Infrared imaging detection of astronomical objects is of great importance in astronomy and astrophysics.A high-performance infrared detector is crucial to infrared astronomical observations.The performance test and evaluation of an infrared focal plane array (IRFPA) detector in this article are significant for infrared astronomy.According to the particularity of astronomical application,we give test principles and methods of performance parameters such as gain,readout noise,linearity and dark current,by analyzing characteristics of an IRFPA detector.A laboratorial testing platform is established to realize the performance test of an IRFPA detector with a wavelength range of 1~14μm.We have completed a performance test of an IRFPA detector made of a domestic 3~5 μm HgCdTe chip.The test results reveal that the linearity of the IRFPA detector is pretty good,over 99.9%.Due to the relatively large readout noise and dark current,there is still a wide gap between the IRFPA detector and high-performance astronomical detectors abroad.
What problem does this paper attempt to address?