Development and application of the cell bioelectrochemical noise analyzer

Xun HUANG,Yao YAO,Rong-hua SHI,Wei WEI,Dan XIAO,Shou-guo WU
DOI: https://doi.org/10.3969/j.issn.1004-1656.2017.02.003
2017-01-01
Abstract:In this paper,the test method of the bioelectrochemical noises of the cellular ion channel is proposed. The electronic cir-cuits for measurement of the electrode potential and the micro-current are designed and developed. The Prussian blue modified glassy carbon electrode acts as the sensor of potassium ion to measure the switching period of the cellular potassium ion channel. As a result,the two current noise power spectrum peeks of the potassium ion channel of the cervical tumor cell (HeLa )appear in 0. 04Hz and 0. 06Hz respectively,which indicates that the switching period of the HeLa cellular potassium ion channel is between 16s and 25s. The relative standard deviation(RSD)for eleven measurements is less than 10%.