In Situ Monitoring of Chlorophyll Fluorescence Using PAM Fluorometer

HAN Zhi-guo,LEI La-Mei,HAN Bo-ping
DOI: https://doi.org/10.3969/j.issn.1008-8873.2005.03.013
2005-01-01
Abstract:This paper introduced an in situ monitoring technique of chlorophyll fluorescence using portable Pulse-Amplitude Modulation fiuorometer, PAM-2100. With this technique Ft, Fm', Y, rETR, qP, qN or NPQ, PAR, and leaf temperature can be monitored without disturbing plant natural photosynthesis. An in situ monitoring case with Eichhornia crassipes was examined. Chlorophyll fluorescence were measured every 5 min and monitored continuously for 225 min. The results indicated that the changes of chlorophyll fluorescence parameters depended on the variation of environmental PAR. Some parameters such as Ft, rETR, qN and NPQ showed a consistent change with PAR, while Fm', Y and qP changed reversely. This case demonstrated that in situ monitoring technique was reliable, simple, rapid and sensitive. Along with the appearance of new modulated fluorometers suit for in situ monitoring, this technique may be widely used in plant ecology research.