Phase Resolved Polarization Sensitive Optical Coherence Tomography:Imaging and Quantification of Tissue Birefringence

丁志华,任洪武,陈忠平,吴兰
DOI: https://doi.org/10.3969/j.issn.1007-7146.2003.03.002
2003-01-01
Abstract:A fiber-designed phase resolved polarization-sensitive optical coherence tomography (PS-OCT) system was developed. The polarization modulator in this system was introduced in the reference arm rather than the source arm, providing an increased power delivered from light source to sample. As the source is only partially polarized, the light incident on the sample has components of any polarization states. Modulating the polarization states of light at the reference arm, backscattered light that originates from incident light upon sample at four orthogonal polarization states was coherently picked out. Based on angle preservation of Stokes vectors in the PS-OCT system, Stokes parameters of backscattered light measured at the detection arm were used to determine the phase retardation of birefringence samples as a function of depth. Using this PS-OCT system, birefringence of ligament and cartilage are confirmed. Investigation on birefringence alternation of ligament under different conditions is conducted. Ligament relaxation induces significant lose of birefringence, while the residual effect of stretching on ligament is not evident.
What problem does this paper attempt to address?