The Inlfuence of Carbon Fibre Materials on Dose Veriifcation in Radiotherapy Regimen for Nasopharyngeal Carcinoma

Fei-yue SHI,Wei QIN,Huan-yu ZHAO,Qiang LIN,Jin-fei CHEN
DOI: https://doi.org/10.3969/j.issn.1674-1633.2016.01.032
2016-01-01
Abstract:Objective To investigate the influence of carbon fibre materials (treatment couch top and immobilization baseplate) on the results of dose verification by using Varian Portal Dosimetry (PD) system for treatment regimen of nasopharyngeal carcinoma.Methods CT images and structures of a patient with nasopharyngeal carcinoma, which were provided by the 2015 VARIAN Eclipse Treatment Planning Competition, were adopted to design the treatment regimen and to generate corresponding PD dose veriifcation plans. A Varian Clinac IX linac was employed to deliver the veriifcation plans for the following three cases: A. There were no carbon ifbre materials between the linac head and the amorphous silicon electronic portal imager of EPID; B. The Varian image-guided radiation therapy (IGRT) carbon ifbre couch top was put between the linac head and the portal imager; C. Both the Varian IGRT couch top and the ORFIT AIO carbon fibre immobilization baseplate were put between the linac head the portal imager. The PD software module was used to analyze the two-dimensional dose distributions for each ifeld. Three sets of Gamma pass rate data were acquired, with each set consisting of 18 Gamma pass rate values.Results Statistical analysis was applied to the three sets of Gamma pass rate data. The statistical results of mean± standard deviation were as follows: Case A (94.31%±1.75%), Case B (94.35%±1.81%), Case C (91.58%±2.91%). The results of t test show that, there is no statistical difference between Gamma pass rate data of Case A and Case B(P>0.05), whereas there is a statistical difference between those of Case A and Case C(P=0.001).Conclusion For treatment regimens of nasopharyngeal carcinoma, the carbon ifbre couch top has little inlfuence on the PD dose veriifcation. Yet the combination of carbon ifbre couch top and baseplate has signiifcant inlfuence on the dose veriifcation, resulting in obvious decrease in Gamma pass rate values and thus worsening of the veriifcation results.
What problem does this paper attempt to address?