Determination of lead and cadmium in blood and urine by microwave digestion-graphite furnace atomic absorption spectrometry

Yu-bo LIU,Yi-bing ZHOU,Jiang HUANG,Jie WANG,Jia-lin DAI,Zheng REN,Shun-qin CHEN,Bing XIA
2017-01-01
Abstract:Objective The study armed to establish a method for determination oflead and cadmium contents in blood and urineby microwave digestion-graphite furnace atomic absorption spectrometry,so as to provide a reference for clinical diagnosis and treatment of lead and cadmium poisoning.Methods 0.50 mL of blood and urine samples were subjected to microwave digestion for 5min in HNO3/H2O2 digestion solution.Ammonium sulfate and ammonium dihydrogen phosphate mixture was used as matrix modifier for graphite furnace atomic absorption spectrometry to measure the lead and cadmium contents in the digestion solution.Results The lead and cadmium contents showed good linearity within the ranges of 0.0-160μg/L and 0.0-16.0μg/L,respectively (the correlation coefficient was 0.9993 for lead and 0.9987 for cadmium).The detection limits (3s/b)of the method were0.36μg/L and 3.24μg/L,respectively.Average recovery rate of thesamples ranged between 97.8%and 103.0%,and the relative standard deviation (RSD) (n=6) was 1.76%-6.81%.Conclusion The method isfastand meets the requirements for precision and accuracy,providing a simple and accurate determination method for clinical diagnosis of lead and cadmium poisoning.