The Comparison of Various Drainage Methods in Minimally Invasive Robotic Assisted Heart Surgery under Cardiopulmonary Bypass

Yun Zhao,Yue Cheng,Kejian Hu,Lai Wei,Jinqiang Shen,Chunsheng Wang
DOI: https://doi.org/10.3969/j.issn.1674-1242.2014.02.007
2014-01-01
Abstract:Objective To compare and analyze the characteristics and management of different methods of venous drainage and auxiliary drainage during robotic assisted on -pump minimally invasive cardiac surgery. Methods 24 cases of patients underwent on-pump cardiac surgical procedures with peripheral extracorporeal cir-culation established by cannula of femoral arterial and vein using Da Vinci S surgical system. The auxiliary drainage methods were divided into 3 groups,centrifugal pump auxiliary drainage(D1 group,n=12),vacuum-assisted drainage(D2 group,n=7),and internal jugular vein cannula auxiliary drainage(D3 group,n=5). Result The average venous drainage and the average venous drainage per kilogram were similar in three groups without sta-tistical difference. Steady transit operation flow,stable hemodynamics,normal monitoring indicators and acceptable surgical field exposure were observed during pump cardiac surgery of three groups . The mean peripheral extracor-poreal and aortic cross-clamping time were 139. 7 ± 35. 5mins and 84. 2 ± 28. 8mins,respectively. Conclusion On the basis of femoral venous drainage,perfusionists should apply different drainage methods and make cor-responding strategies,according to different surgical procedures and surgeons to ensure adequate and sufficient drainage and perfusion. This is one of the most important factors of successful robotic assisted on-pump cardi-ac surgeries.
What problem does this paper attempt to address?