Research of three-dimensional ultrasound system based on infra positioning system

Han ZHOU,Bing LI,Jun-ting CHEN,Xi-xu ZHU,Jun-shu SHEN,Yun GE
DOI: https://doi.org/10.13339/j.cnki.sglc.20160303.002
2016-01-01
Abstract:Objective To reconstruct three-dimensional ultrasound system by the real-time infra positioning system and collection of two-dimensional ultrasound images, and timely track the target area by breathing and respiration movement. Methods Three-dimensional ultrasound data was constructed based on two-dimensional ultrasound probe installed with three infrared positioning balls(non-equilateral), which tracked the position of images, real-time mapping the images into the three dimensional space. The cube model was used as phantom, and whole experimental environment was conducted in sink. Results Three-dimensional ultrasound system was rapidly reconstructed by the proposed method, and the reconstruction of triangular pyramid was fitted for real phantom with high reconstruction precision. Conclusion It is demonstrated that the real-time tracking technique can realize calibration and reconstruction of three-dimensional ultrasound rapidly and easily, and it also possesses high reconstruction precision.
What problem does this paper attempt to address?