Reticulocyte Analysis by FCM and Its Application in Anaemia
ZANG Wan,WU Li-juan,CHEN Wei
DOI: https://doi.org/10.3969/j.issn.1671-8348.2008.03.011
2008-01-01
Abstract:Objective To establish a method for analyzing reticulacytes by Flow Cytometer and research its clinical applicated values in diagnosing anaemia.MethodsThiazole orange was used as a fluorochrome label to stain reticulocytes by combinding to RNA in reticulcyte plasma.The percent,absolute counting and fluorescence intensity,which images the amount of RNA in reticulcyte plasma,were detected by Flow Cytomter.ResultsIn health control Group,the percent,absolute counting and fluorescence intensity of reticulocyte was(0.011 4±0.004 0),(49.75±63.27) and(22.56±34.57)×109/L,respectively.In hemolytic anemia group,it was(4.218 0±0.265 1),(1 795.25±549.16) and(43.15±48.05)×109/L,respectively.In iron deficiency anemia group,it was(0.027 3±0.013 7),(86.57±67.28) and(23.09±21.00)×109/L,respectively.In blood loss anemia group,it was(3.978 2±1.089 6),(2 048.01±604.52) and(39.46±18.12)×109/L,respectively.In chronic renal insufficiency group,it was(0.005±0.002 5),(21.59±5.64) and(18.25±9.14)×109/L,respectively.In tumour chems group,it was(0.020 3±0.229),(24.90±43.35) and(14.27±8.13)×109/L,respectively.Compared every anemia group with health control group,the results of percent and absolute counting were obviously different.ConclusionFlow Cytometry analyzing reticulocytes is simple,fast and accurate,and has a good reproducibility.It could use in the diagnosis and curative effect monitoring for various kinds anemia.