Determination of mercury in whole blood by inductively coupled plasma-mass spectrometry

Junxia LIU,Xin ZHOU,Chonghuai YAN
DOI: https://doi.org/10.3969/j.issn.1673-8640.2015.12.016
IF: 1.865
2015-01-01
Laboratory Medicine
Abstract:Objective To study the feasibility of inductively coupled plasma-mass spectrometry (ICP-MS)for whole blood mercury determination.Methods Using hydrochloric acid as sample matrix,samples were digested with microwave digestion instrument,and ICP-MS was used to determine mercury in whole blood.The lowest detection limit, functional sensitivity,range,recovery,accuracy and precision of ICP-MS were evaluated.Results The detection limit of this method was 4.525 ×1 0 -2 ng/mL,the functional sensitivity was 1 .495 ng/mL,the analytical measurement range was 0.1 8-1 3.83 ng/mL,and the recovery rate was 92%-1 03%.The average values of low-level [(20 ±5)ng/mL ] and high-level [(50 ± 8)ng/mL]quality control materials for 20 d were 22.22 ng/mL and 51 .39 ng/mL,and between-run precisions were 6% and 5%.Clinical random mixed whole blood,low and high whole blood quality control materials for 20 parts each were detected within 1 d,and the within-run precisions were 4.8%,4.8% and 4.4%. Conclusions ICP-MS for the determination of whole blood mercury meets the clinical requirements.
What problem does this paper attempt to address?