Evaluation of Perceval Sorin Prothesis by Echocardiography

Bo Zhang,Ming Chen
DOI: https://doi.org/10.3969/j.issn.1002-0101.2010.04.010
2010-01-01
Abstract:Objective To explore the value of echocardiography in a clinic trial of Perceval Sorin prothesis.Methods Retrospective study was made in 15 elderly patients with aortic valve disease who received replacement of Perceval Sorin prosthesis.TTE and TEE were performed pre operation,one week and one month post operation,which were applied to detect the flow and function of the aortic valve or prothesis.Results Moderate-severe aortic stenosis was detected in 15 patients before aortic valve replacement.The mean gradient across the prosthesis was 12.57±1.99 mm Hg post surgery.The area of the prosthesis was 1.37±0.28 cm~2.No leakage was found and only one case with mild regurgitation in 15 patients.Conclusions After the replacement of Perceval Sorin prothesis,the pressure gradient of aortic valve decreased significantly,the area of aortic valve became larger than before.Echocardiography could provide an reliable tool in assessment of function and efficiency of the prosthesis,with valuable information in patients with Perceval Sorin prothesis.
What problem does this paper attempt to address?