Influence of imaging parameters on the detection of gadolinium concentration in artificial cerebrospinal fluid in fluid-attenuated in version recovery MR Imaging at low field strength

De-sheng ZHANG,Xiang-sheng XIAO,Shi-yuan LIU,Xiao-feng TAO,Shu-jin WANG,Wei-jian SONG
DOI: https://doi.org/10.3969/j.issn.1673-6966.2013.08.011
2013-01-01
Abstract:Objective To evaluate the effects of different fluid-attenuated inversion recovery (FLAIR) sequences on the detection of gadolinium concentration in artificial cerebrospinal fluid (CSF) at low field MRI. Methods Brain parenchyma and different enhancement in CSF were simulated with fresh hard-boiled egg whites and mixtures of artificial CSF and gadolinium diethylenetriamine pentaacetic acid (Gd-DTPA), with the concentration ranging from 0 mmol/L to 8 mmol/L. Turbo FLAIR MR sequences at different values of repetition time (TR), inversion time(TI), and effective echo time (TEeff) were performed at 0.35-T MR scanner. The influence of the imaging parameters on the detection of gadolinium concentration in the artificial CSF was defined. Results At a fixed TEeff of 150 ms, when the TR were increased from 5000 ms to 20000 ms (thus the TI increased from 1600 ms to 2800 ms), the detectability for gadolinium concentration was about 0.0078 mmol/L and unchanged. At a fixed TR of 7079 ms and TI of 2000 ms, the discriminable gadolinium concentration was decreased from 0.0625 mmol/L to 0.002 mmol/L as the TEeff increased from 20 ms to 280 ms(P<0.0001). Conclusion When the TR is longer than 5000 ms in the T2-weighted FLAIR sequence at 0.35-T, the discriminable gadolinium concentration in CSF decreases with the increase of TEeff and is nearly unaffected by TR (TI) used.