Reproducibility study of Fibroscan for liver controlled attenuator parameters measurement

Guang-yu CHEN,Qin PAN,Feng SHEN,Jia-fa LU,Rui-nan ZHANG,Chan-yan ZHU,Wan-lu SUN,Jing ZENG,Jia-yu WU,J FAN
DOI: https://doi.org/10.3969/j.issn.1008-1704.2015.11.005
2015-01-01
Abstract:Objective To evaluate the repeatability of Fibroscan in detecting liver controlled attenuator parameter (CAP)between different operators,and to investigate its diagnosis agreement for fatty liver.Methods Seven subjects (healthy volunteers,fatty liver patients and other diseases patients)were enrolled for CAP measurement by five operators with different levels of operative experiences.The repeatability and diagnosis agreement of different operators were assessed by intraclass correlation coefficient (ICC)and Kendall's coefficient.Results The ICC values of five operators,experienced operators and less experienced operators were 0.959 (95% CI 0.880,0.992),0.913 and 0.876 (P =0.36),respectively, which displayed a favorable repeatability.The Kendall's coefficient was 0.715 in assessing fatty liver grade by CAP value, which revealed a satisfactory consistency (P =0.002).Conclusion Fibroscan in CAP measurement showed not only better repeatability but also satisfactory consistency among operators with varying experiences.