Abstract:Objective To observe the accuracy,effectiveness and safety of radiofrequency catheter ablation (RFCA) of sustained atrial fibrllation (AF) by using the domestic ColumbusTM cardiac three dimensional electromagnetic mapping system (ColumbusTMsystem).Methods The study population was composed of 30 patients,26 male,mean age 41-75(59.0±8.8) years,with sustained AF for 1-12 months.All the 30 patients underwent circumferential pulmonary vein isolation (CPVI) by using ColumbusTMsystem.In some patients,additional linear ablation of left atrial roof line and/or tricuspid annulus isthmus hne was added.Results The endpoints of CPVI achieved in all the 30 patients.The 12th and 13th patients underwent a linear ablation of the left atrial roof line or tricuspid annulus isthmus line,and a bidirectional conduction block achieved in all patients.During the procedures,the ColumbusTM system had an accurate localization,and had no displacement.There were no serious ablation-related complications during and after the operation procedures.During a period of 12 months of follow-up,23 patients (76.7%) had no AF recurrence,and AF or atypical atrial flutter recurred in 6 and 1 patients.One of the 7 patients underwent a re-ablation procedure,and the patient had no recurrence during 6 months of follow-up.Conclusion Domestic ColumbasTM system was accurate,effective and safe in RFCA of sustained AF.