Preliminary experience in extraction of pacemaker leads with Evolution mechanical dilator sheath

Xiaohong Li,Wenqing Ji,Wei Xu
DOI: https://doi.org/10.3760/cma.j.issn.1007-6638.2014.06.009
2014-01-01
Abstract:Objective Observe the safety and effectiveness of lead extraction with the Evolution mechanical dilator sheath.Methods From June 2013 to June 2014,patients with cardiac device infection undergoing lead extraction with different methods were enrolled in this study.The procedures were performed with mechanical sheaths or Evolution system to compare the safety and effectiveness.Results Twenty-four patiens were enrolled.47 leads (16 in atrial,31 in ventricular)were extracted and 14 leads were removed with Evolution mechanical dilator sheath.Compared with mechanical sheaths,Evolution system reduced the operation time [(2.1±1.2) h vs.(1.5±0.9) h,P =0.04] and X-ray time[(16.5±6.2) min vs.(8.7±4.2) min,P=0.04].Complete procedural success with Evolution system was achieved in 14 leads.All of the patients discharged uneventfully without a major complication.Conclusion Our experience confirms that the mechanical technique with Evolution system reduce the operation time and X-ray time,the technique is an effective and safety method for lead extraction.