The Development of Gasless Laparoscopic Devices and Gasless Laparoscopic Surgery

王秋生,冷希圣,李恩宽,岳勇,吕杰
DOI: https://doi.org/10.3760/j:issn:0529-5815.2001.04.008
2001-01-01
Abstract:Objective To study the clinical safety and feasibility of gaslesslaparoscopic surgery with our homemade gasless laparoscopic devices. Methods Five generations of gasless laparoscopic device have been designed and manufactured independently since 1993, and 12 sorts and 66 cases of gasless laparoscopic procedure have been developed in the clinic after experimental research. Results Except 5 cases were converted to open surgery(conversion rate 7.5%), the other 61 cases of gasless laparoscopic cholecystectomy(41), inguinal herniorrhaphy(2), abdomino-perineal resection of rectal carcinoma(9), sigmoidectomy(3), left hemi-colectomy(1), hand-assisted sigmoidoproctectomy(1), hand-assisted right hemi-colectomy(1), partial excision of cecum(1), sub-total gastrectomy(1), and abdominal plasty(1) were successfully performed with no mobidity and mortality. Conclusions gasless laparoscopic surgery is not only safe and feasible but also reflects complementary advantages of modern laparoscopic surgery and traditional open surgery. Besides its supplementary value for the patients with cardiopulmonary diseases contraindicated to the pneumoperitoneum, it also possesses distinctive value to both gasless laparoscopic gastrointestinal surgery with an inherent incision or an incision for specimen retrieval and hand-assisted gasless laparoscopic hepatobiliary and pancreatic surgery.
What problem does this paper attempt to address?