A feasibility study of MR thermometry in vertebra tumor thermotherapy

Shuo CHEN,Guang-ze SONG,Fei-yu CHEN,Xin-wei SHI,Bing-yao CHEN,Xing WEI,Kui YING
DOI: https://doi.org/10.3969/j.issn.2095-252X.2015.11.015
2015-01-01
Abstract:Objective To analyze dififculties of temperature measuring by MR thermometry in different types of tissues near vertebra.Method Regions of interest ( ROI ) were segmented into several parts based on clinical needs and T1 / T2 maps were acquired in a 3T Phillips scanner. Proton Resonance Frequency ( PRF ) and spectrum estimation were used separately to measure the temperature in water-domain and water-fat mixed tissues, with ifeld drift correction using muscles as background tissues.Results Eighty-three hundred pixels were chosen in each ROI for PRF method. The mean error value of the spinal cord and intervertebral disk was below 0.2℃, and the standard deviation was below 1.5℃. The mean error value of the aorta and vena cava was 0.9℃-1.8℃, while the standard deviation was below 1.7℃. The mean error value of the vertebra was around 0.9℃, while the standard deviation was near 12℃. The standard deviation of the vertebra water-fat mixed region was still above 12℃ by using spectrum estimation method. Conclusions MR thermometry nas good performances in regions of water-domain and less blood lfow, while the blood lfow in aorta and vena caca can induce disturbances in temperature measuring. Susceptibility caused by cancellous bone can also bring errors into temperature results.
What problem does this paper attempt to address?