Effects of energy level of electric knife on the recurrent laryngeal nerve in thyroid surgery

刘蟩薇,樊友本,郑起,邓先兆,康杰,伍波,杨治力,周鸣
DOI: https://doi.org/10.3969/j.issn.1005-6483.2015.02.016
2015-01-01
Abstract:Objective To study the influence of energy level of electric knife on the recurrent la-ryngeal nerve in a porcine animal model and discuss the safe energy range of high-frequency electric knife in surgical operation.Methods The experiment was set up in 12 pigs.The distance from the head of elec-tric knife to the nerve was one millimeter and the action time was three seconds.The animals were random-ly divided into three groups according to the energy level(45 W,90 W,and 135 W).Preoperative and postoperative electromyogram of laryngeal muscle were recorded and analyzed by a laryngeal nerve detec-tor.Nerve tissues were extracted to make pathological sections for histological observation.Results The nerve detector indicated that the difference in amplitude between group 2(90 W)and group 3(135 W)was significant(P <0.05).No significant difference was observed in group 1(45 W).Histological study shows that obvious injury in group 3(135 W).Conclusion In thyroid surgery,the safe energy range of electric knife for recurrent laryngeal nerve should be less than 90 W.
What problem does this paper attempt to address?